ZEISS i.Profiler plus Accurate analysis for better vision, day and night.
The i.Profiler plus by ZEISS is a 4-in-1 compact system that combines an ocular wavefront aberrometer, autorefractometer, ATLAS corneal topographer and keratometer in a single device. The most accurate lenses can only be prescribed when you have all the relevant information about a patient’s eyes. Let the ZEISS i.Profiler plus provide you with a detailed visual profile so you can answer your patients’ visual needs.

Assist with finding and explaining the best possible solution for each patient.
- ZEISS i.Profiler plus features a high-resolution wavefront measurement and corneal topography, providing you with all the necessary data to fully evaluate the refractive status of your patient’s eyes.
- The analysis mode enables you to visualise the impact of different aberrations on your patient’s vision, including up to 7 Zernike aberrations.
- It enables prescribing of ZEISS Lenses with i.Scription Technology, using the generated ‘fingerprint’ measurement of each eye.
- Connectivity to the optional ATLAS Review for ZEISS i.Profiler plus software enables advanced keratometry and topography, enabling contact lens fitting and more.

Gives a comprehensive overview of a patient’s visual performance and refractive conditions.
- Ability to provide data to support identification and analysis of ocular pathologies or abnormalities.
- Supports the user in the identification of ideal candidates for ZEISS i.Scription lenses.
- Provides full analysis of the eye's refractive power distribution across the entire pupil aperture, not just the central zone.

Fast, comprehensive and accurate.
- Fully-automated measurement procedure.
- Touch screen control enables all measurements of both eyes in 60 seconds.
- High-resolution Hartmann-Shack wavefront sensor measuring 1500 points across a 7-millimeter pupil aperture.
- User-friendly, easy-to-use functions enable capture, evaluation, presentation and analysis of data.
- 50% reduction in chair time for subjective refraction, due to accurate starting data generated by ZEISS i.Profiler plus. 1

Access to i.Scription® Technology by ZEISS.
- The 7 visual optics analysis parameters summarize and evaluate the interactions of measurements obtained from the wavefront analyser, the detected pupil and the corneal topographer.
- This provides an automatic and complete assessment of the eye optics, enabling a fast and accurate subjective refraction, and the identification of ‘i.Scription ready’ patients.
- With ZEISS i.Scription Technology, you can offer each patient a unique lens solution, improving their vision day and night.
- The benefits of i.Scription® Technology by ZEISS can be simulated and demonstrated to the patient using the point spread function graphs and comparison (with and without i.Scription).

Outstanding connectivity: incorporate it into your digital workstream.
- Ability to remotely access the ZEISS i.Profiler plus via VISUCONSULT 500 and demonstrate the patient’s ZEISS i.Scription analysis on the iPad® 2.
- The data captured from the ZEISS i.Profiler plus is saved on the VISUCONSULT 500 data management system, allowing for future consultation use, ordering and archiving, without any need for further measurements.
- It integrates with compatible PMS systems for a smooth and efficient workflow.3
ZEISS i.Profiler plus
Technical specifications
Technical Data Wavefront
|
|
---|---|
Measuring range, cylinder:
|
0 D to +8 D
|
Measuring range, sphere:
|
-20 D to +20 D (3.5 mm pupils)
|
-15 D to +15 D (5.5 mm pupils)
|
|
Axis:
|
0º – 180º
|
Measuring surface:
|
2.0 mm to 7.0 mm (three zones)
|
No. of measuring points:
|
up to 1500
|
Method:
|
Hartmann-Shack
|
Reference wavelength1:
|
555 nm according to ISO 24157
|
Technical Data Corneal Topography
|
|
---|---|
No. of rings:
|
22 (18 complete rings)
|
No. of measuring points:
|
3,425
|
Detected corneal surface at 42.125 D:
|
dia. 0.75 mm to 9.4 mm
|
Dioptres:
|
measuring range 25 to 65 D
|
Accuracy:
|
± 0.05 D (±0.01 mm)
|
Reproducibility:
|
± 0.10 D (±0.02 mm)
|
Type A:
|
according to ISO 19980
|
1 Reference wavelength for the interpretation of refractive errors (referring to maximum luminosity function V(λ) of the human eye in daylight).
Line voltage
|
100-240 V AC ± 10%, 50…60 Hz
|
---|---|
Power consumption
|
≤ 200 VA
|
Dimensions (W x H x D)
|
345 x 555 x 525 mm
|
Weight
|
30 kg
|
Interfaces
|
VGA, 3x USB, RS232, 2x LAN
|
Printer
|
Thermal, integrated
|
Display
|
12" Color Touch LCD
|
-